Specifications
Edan SE-1200 Express Basic 12-Channel ECG Specifications
- Weight: 11 lb (5 kg) excluding recorder paper and battery.
- Dimensions: 16.5" x 13" x 4.7" (420 mm x 330 mm x 120 mm).
- Display: 8" (800 x 600 multicolor LCD screen).
- Power Supply:
Mains Supply:
Operating Voltage: 100 to 240 V.
Operating Frequency: 50/60 Hz.
Input Current: 0.9 to 0.4 A.
Internal Li-ion Battery Pack:
Rated Voltage: 14.8 V.
Rated Capacity: 2500/5000 mAh (1.5/2.5 hours continuous printing, 300/450 ECG reports).
Necessary Charge Time: 3/6 hours.
Recorder: Thermal dot-matrix recorder.
8 dots per mm / 200 dots per inch (amplitude axes).
40 dots per mm / 1000 dots per inch (time axes, at 25 mm/s).
Recorder Paper: SE-12P - Z-Fold Thermal Paper (210 mm x 295 mm x 100 pages).
Paper Speed: 5 mm/s, 6.25 mm/s, 10 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s.
External Printer: HP1010/1510, HP M401, HP 1020/1020PLUS/1106, HP 2010/1050/2000, HP 2015/2035, HP 1525.
HR Range: 30 to 300 BPM.
Accuracy: ± 1 BPM.
- ECG Unit:
Leads: 12 standard.
Acquisition Mode: 12 leads simultaneously.
A/D Converter: 24 bits.
Resolution: 2.52 µV/LSB.
Time Constant: ≥ 3.2 s.
Frequency Response: 0.01 ~ 300 Hz (-3 db).
Gain: 1.25, 2.5, 5, 10, 20, 10/5 mm/mV, AGC.
Input Impedance: ≥ 100 MΩ (10 Hz).
Input Circuit Current: ≤ 0.01 µA.
Input Voltage Range: ≤ ± 5 mVpp.
Calibration Voltage: 1 mV ± 2%.
CMRR:
AC On: ≥ 140 dB.
AC Off: ≥ 123 dB.
Sampling Frequency: 16000 Hz.
Amplitude: ± 750 µV to ± 700 mV.
Width: 50 µs to 2.0 ms.
AC: 50/60 Hz.
DFT: 0.01 Hz / 0.05 Hz / 0.15 Hz / 0.25 Hz / 0.32 Hz / 0.5 Hz / 0.67 Hz.
EMG: Off / 25 Hz / 35 Hz / 45 Hz.
LOWPASS: 300 Hz / 270 Hz / 150 Hz / 100 Hz / 75 Hz.
Report Format: PDF, XML, DICOM, FDA-SCP.
Data Transmission: Wi-Fi, Ethernet, RS232.
SE-1515 Data Management System, bi-directional communication.
HIS Connection: DICOM Worklist / DICOM Storage / HL7 / GDT.
Transmitting Frequency / Frequency Band: 2400 to 2497 MHz.
Modulation Type: DSSS, CCK, OFDM.
Transmitting Power / Effective Radiated Power: 6 to 17 dBm.